Determination of Irrigation Flowrate During Flexible Ureteroscopy: Methods for Calculation Using Renal Pelvis Pressure.
Document Type
Article
Publication Date
8-16-2022
Publication Title
Journal of endourology
Abstract
BACKGROUND: Proper control of irrigation flowrate during ureteroscopy is important to manage thermal and pressure risks. This task is challenging because flowrate is not directly measured by commercially available ureteroscopic or fluid management systems. However, flowrate can be calculated using a hydrodynamic relationship based on measurable values during ureteroscopy. Objectives of this in vitro study were to 1) calculate inflow resistance for different working channel conditions and then using these values 2) calculate irrigation flowrate and determine its accuracy across a range of renal pelvis pressures.
MATERIALS AND METHODS: A 16 Liter container was filled with deionized water and connected by irrigation tubing to a 9.6Fr single-use ureteroscope. Inflow resistance was determined by plotting flowrate (mass of fluid collected from ureteroscope tip in 60 seconds) versus irrigation pressure (range 0-200 cmH2O). Next, the tip of the ureteroscope was inserted into the renal pelvis of a silicone kidney-ureter model and renal pelvis pressure was measured. In conjunction with the previously determined inflow resistance and known irrigation pressure values, flowrate was calculated and compared to experimentally measured values. All trials were performed in triplicate for working channel conditions: empty, 200µm laser fiber, 365µm laser fiber, and 1.9Fr stone basket.
RESULTS: Flowrate was linearly dependent on irrigation pressure for each working channel condition. Inflow resistance was determined to be 5.0 cmH2O/(ml/min) with the 200µm laser fiber in the working channel and calculated flowrates were within 1 ml/min of measured flowrates. Similar results were seen with a 365µm laser fiber, and 1.9Fr basket.
CONCLUSIONS: Utilizing renal pelvis pressure measurements, flowrate was accurately calculated across a range of working channel conditions and irrigation pressures. Incorporation of this methodology into future ureteroscopic systems that measure intrarenal pressure, could provide a real-time readout of flowrate for the urologist and thereby enhance safety and efficiency of laser lithotripsy.
Volume
Online ahead of print.
Recommended Citation
Rezakahn Khajeh N, Hall TL, Ghani KR, Roberts WW. Determination of irrigation flowrate during flexible ureteroscopy: methods for calculation using renal pelvis pressure. J Endourol. 2022 Aug 16. doi: 10.1089/end.2022.0039. Epub ahead of print. PMID: 35974664.
DOI
10.1089/end.2022.0039
ISSN
1557-900X
PubMed ID
35974664