The first prototype of spot-scanning proton arc treatment delivery.
Document Type
Article
Publication Date
8-1-2019
Publication Title
Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology
Abstract
PURPOSE: We report the first prototype of spot-scanning arc treatment (SPArc) delivery on a clinical proton beam therapy machine and evaluate its delivery accuracy and efficiency.
METHODS AND MATERIALS: A new module called Proton Dynamic Arc Delivery (PDAD) was developed to allow simultaneously delivering spot-scanning proton beam treatments while rotating the gantry on an IBA Proteus®One proton machine. A series of measurements was performed to validate the basic beam characteristics. Subsequently, patient specific quality assurance (QA) of a brain SPArc plan was performed. Total SPArc treatment delivery time was also recorded and compared to the clinically delivered intensity modulated proton therapy (IMPT) treatment time. Finally, the log file of the SPArc plan was analyzed and processed to reconstruct the actual delivered dose.
RESULTS: All the basic beam characteristics were confirmed in the PDAD mode, similar as those measured using fixed gantry deliveries in clinical mode. The brain SPArc plan with similar or superior plan quality was delivered in 4 mins compared to total 11 mins for the clinical treatment of the three-field IMPT plan. The patient QA result showed a good agreement between the measured and calculated dose distributions with the gamma index of 98.6% (3%/3 mm). The analysis of the log file confirmed the accuracy of the SPArc plan delivery, with the gamma index of 98.3% (1%/1 mm) between reconstructed and the planned doses.
CONCLUSION: The first prototype of dynamic proton arc delivery on a clinical proton therapy system was successfully performed. The measurements and simulations demonstrated the feasibility of SPArc treatment within the clinical requirements.
Volume
137
First Page
130
Last Page
136
Recommended Citation
Li X, Liu G, Janssens G, De Wilde O, Bossier V, Lerot X, Pouppez A, Yan D, Stevens C, Kabolizadeh P, Ding X. The first prototype of spot-scanning proton arc treatment delivery. Radiother Oncol. 2019 Aug;137:130-136. doi: 10.1016/j.radonc.2019.04.032. Epub 2019 May 14. PMID: 31100606.
DOI
10.1016/j.radonc.2019.04.032
ISSN
1879-0887
PubMed ID
31100606