"Quantifying pulmonary perfusion from noncontrast computed tomography." by Edward Castillo, Girish Balachandran Nair et al.
 

Quantifying pulmonary perfusion from noncontrast computed tomography.

Edward Castillo, Beaumont Health
Girish Balachandran Nair, Beaumont Health
Danielle Turner-Lawrence, Beaumont Health
Nicholas Myziuk, Beaumont Health
Scott Emerson, Beaumont Health
Sayf Al-Katib, Beaumont Health
Sarah Westergaard
Richard Castillo
Yevgeniy Vinogradskiy
Thomas Quinn, Beaumont Health
Thomas Guerrero, Beaumont Health
Craig Stevens, Beaumont Health

Abstract

PURPOSE: Computed tomography (CT)-derived ventilation methods compute respiratory induced volume changes as a surrogate for pulmonary ventilation. Currently, there are no known methods to derive perfusion information from noncontrast CT. We introduce a novel CT-Perfusion (CT-P) method for computing the magnitude mass changes apparent on dynamic noncontrast CT as a surrogate for pulmonary perfusion.

METHODS: CT-Perfusion is based on a mass conservation model which describes the unknown mass change as a linear combination of spatially corresponding inhale and exhale HU estimated voxel densities. CT-P requires a deformable image registration (DIR) between the inhale/exhale lung CT pair, a preprocessing lung volume segmentation, and an estimate for the Jacobian of the DIR transformation. Given this information, the CT-P image, which provides the magnitude mass change for each voxel within the lung volume, is formulated as the solution to a constrained linear least squares problem defined by a series of subregional mean magnitude mass change measurements. Similar to previous robust CT-ventilation methods, the amount of uncertainty in a subregional sample mean measurement is related to measurement resolution and can be characterized with respect to a tolerance parameter

RESULTS: The median correlations between CT-P and SPECT-P taken over all 30 test cases ranged between 0.49 and 0.57 across the parameter sweep. For the optimal tolerance τ = 0.0385, the CT-P and SPECT-P correlations across all 30 test cases ranged between 0.02 and 0.82. A one-sample sign test was applied separately to the PE and lung cancer cohorts. A low Spearmen correlation of 15% was set as the null median value and two-sided alternative was tested. The PE patients showed a median correlation of 0.57 (IQR = 0.305). One-sample sign test was statistically significant with 96.5 % confidence interval: 0.20-0.63, P < 0.00001. Lung cancer patients had a median correlation of 0.57(IQR = 0.230). Again, a one-sample sign test for median was statistically significant with 96.5 percent confidence interval: 0.45-0.71, P < 0.00001.

CONCLUSION: CT-Perfusion is the first mechanistic model designed to quantify magnitude blood mass changes on noncontrast dynamic CT as a surrogate for pulmonary perfusion. While the reported correlations with SPECT-P are promising, further investigation is required to determine the optimal CT acquisition protocol and numerical method implementation for CT-P imaging.