Development of a standalone delivery sequence model for proton arc therapy.

Document Type

Article

Publication Date

4-2024

Publication Title

Medical physics

Abstract

BACKGROUND: Spot-scanning proton arc (SPArc) has been drawing significant interests in recent years because of its capability of continuous proton irradiation during the gantry rotation. Previous studies demonstrated SPArc plans were delivered on a prototype of the DynamicARC solution, IBA ProteusONE.

PURPOSE: We built a novel delivery sequence model through an independent experimental approach: the first SPArc delivery sequence model (DSM

METHODS: SPArc test plans were generated and delivered on a prototype of the DynamicARC solution, IBA ProteusONE. An independent gantry inclinometer and the machine logfiles were used to derive the DSM

RESULTS: The difference between the logfile and DSM

CONCLUSIONS: The first model of dynamic arc therapy is established in this study through an independent experimental approach using logfiles and measurements which allows clinical users and investigators to simulate the dynamic treatment delivery and assess the daily treatment throughput improvement.

DOI

10.1002/mp.16879

ISSN

2473-4209

PubMed ID

38064634

Share

COinS