Understand the FDA-cleared fentanyl testing: A clinical evaluation of the SEFRIA fentanyl immunoassay.

Document Type

Article

Publication Date

6-1-2024

Publication Title

Drug and alcohol dependence

Abstract

BACKGROUND: Screening for fentanyl has been adopted by many clinical laboratories to detect illicit drug use and monitor medication adherence. However, compared to other urine drug testing, fentanyl screening assays are relatively new and therefore their clinical performances are largely unknown. This study extensively evaluated the clinical performance, positive cutoff, and interference profile of SEFRIA fentanyl immunoassay in real patient settings.

METHODS: The FDA-cleared cutoff of 1.0 was verified with 21 urine samples with low or undetectable levels of fentanyl. After assay implementation, all screened-positive samples were confirmed by liquid chromatography-tandem mass spectrometry. A new cutoff was derived from the numeric values of the false positive (FP) screening results. The FP rates before and after implementing the new cutoff were compared. Interferences were identified by an untargeted drug analysis and confirmed by spiking experiments.

RESULTS: A total of 3951 screening results were reviewed in the first two months of the assay utilization, 410 were screened-positive, and 157 (38 %) were FP. After a new cutoff of 1.3 was implemented, the FP rate was reduced to 17 % based on 11119 screening results. Trazodone, labetalol, and haloperidol were identified as major interferents, accounting for 56 % of the FP results using the cutoff of 1.3.

CONCLUSION: By applying the new cutoff and including an interference comment to positive screening results, the FP rate was reduced from initial 38 % to 7.5 % (17 % times 56 %).

Volume

259

First Page

111287

DOI

10.1016/j.drugalcdep.2024

ISSN

1879-0046

PubMed ID

38640864

Share

COinS