Minimizing radiographic contrast administration during coronary angiography using a novel contrast reduction system: A multicenter observational study of the DyeVert™ plus contrast reduction system.
Document Type
Article
Publication Date
6-1-2019
Publication Title
Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions
Abstract
Objective: To evaluate contrast media (CM) volume (CMV) saved using the DyeVert™ Plus Contrast Reduction System (DyeVert Plus System, Osprey Medical) in patients undergoing diagnostic coronary angiogram (CAG) and/or percutaneous coronary interventional (PCI) procedures performed with manual injections.
Background: Current guidelines advocate for monitoring and minimization of the total volume of CM in chronic kidney disease (CKD) patients undergoing invasive cardiac procedures. The DyeVert Plus System is an FDA cleared device designed to reduce CMV delivered during angiography and permit real-time CMV monitoring.
Methods: We performed a multicenter, single-arm, observational study. Eligible subjects were ≥ 18 years old with baseline estimated glomerular filtration rate (eGFR) 20-60 mL/min/1.73 m2 . The primary endpoint was % CMV saved over the total procedure. A secondary objective was to evaluate adverse events (AEs) related to DyeVert Plus System or to CM use.
Results: A total of 114 subjects were enrolled at eight centers. Mean age was 72 ± 9 years, 72% were male, and mean body mass index was 29 ± 5. Baseline eGFR was 43 ± 11 mL/min/1.73 m2 . CAG-only was performed in 65% of cases. One hundred and five subjects were evaluable for the primary endpoint. Mean CMV attempted was 112 ± 85 mL (range 22-681) and mean CMV delivered was 67 ± 51 mL (range 12-403), resulting in an overall CMV savings of 40.1 ± 8.8% (95% CI 38.4, 41.8; P < 0.0001) per procedure. Image quality was maintained in all but one case where the system was turned off for one injection. No DyeVert Plus System-related AEs were reported. Acute kidney injury (AKI; defined as serum creatinine rise of >0.3 mg/dL from baseline) was reported in 11 cases with seven occurring in subjects with baseline eGFR < 30 and three AKI events were attributed to CM. AKI rates increased as CMV/eGFR ratios increased.
Conclusions: These data suggest DyeVert Plus System use in CKD patients undergoing CAG and/or PCI results in clinically meaningful CMV savings while maintaining image quality.
Volume
93
Issue
7
First Page
1228
Last Page
1235
Recommended Citation
Gurm HS, Mavromatis K, Bertolet B, Kereiakes DJ, Amin AP, Shah AP, Hanzel GS, Rao S, Thomas JL, Kumar G. Minimizing radiographic contrast administration during coronary angiography using a novel contrast reduction system: A multicenter observational study of the DyeVert™ plus contrast reduction system. Catheter Cardiovasc Interv. 2019 Jun 1;93(7):1228-1235. doi: 10.1002/ccd.27935. Epub 2018 Nov 4. PMID: 30393942.